Health & beauty United States Recall (ongoing)

AVID Medical, Inc. recalls Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH31. and 1 more products

What's being recalled

Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH31.

Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH30.

Recalling firm: AVID Medical, Inc., Toano, VA, United States.

How to identify it

Model/Catalog Number: SOAH31. UDI-DI: 10809160512890. Lot Code: Lot Number: 1689215. Expiration Date: 11/06/2026. Model/Catalog Number: SOAH30. UDI-DI: 10809160520246. Lot Code: Lot Numbers: 1705826, 1705827. Expiration Date: 1/31/2028.

The hazard

A non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications.

What to do

Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.

The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.

Where it was sold

US Nationwide distributor in the state of Arizona.

The official notice

This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.