Skip to content

Health & beauty Canada Health product recall

RotaFlow Console

What's being recalled

Companies: Maquet Cardiopulmonary Gmbh, Kehler Strasse 31, Rastatt, Germany, 76437.

How to identify it

Affected products: RotaFlow Console; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 70104.3292

The hazard

During internal validation of new replacement batteries, MCP identified an issue on the RotaFlow console. This issue can cause the device to stop battery charging after the battery operation mode has been utilized. Further investigation identified a software problem as root cause. Because of this, all devices on the market are considered affected by this issue.

What to do

Contact the manufacturer if you require additional information.

The official notice

This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.

Read the official notice (opens in a new tab)