Stryker Endoscopy’s 1788 4K Camera family is affected: Camera Head C‑Mount with Advanced Imaging Modality (model 1788210105, lot/serial 23H519774); 1788 4K Camera Control Unit with Advanced Imaging Modality (model 1788010000); and 1788 4K Camera Head with Integrated Coupler (model 1788610122; multiple serials — contact manufacturer). The recall was issued after an increase in out‑of‑box failures from bent connection pins, which may also damage the mating connector in the CCU. Contact the manufacturer for additional information.
What's being recalled
Companies: Stryker Endoscopy, A Div. Of Stryker Corp. Dba Stryker Endoscopy, 5900 Optical Court, San Jose, California, United States, 95138.
How to identify it
Affected products: 1788 4K Camera Head, C-Mount With Advanced Imaging Modality; Lot or serial number: 23H519774; Model or catalogue number: 1788210105
Affected products: 1788 4K Camera Control Unit With Advanced Imaging Modality; Lot or serial number: Not applicable; Model or catalogue number: 1788010000
Affected products: 1788 4K Camera Head With Integrated Coupler With Advanced Imaging Modality; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 1788610122
The hazard
Stryker Endoscopy has received an increase in complaints for 1788 out of box failures due to bent connection pins. The bent pin may also damage the mating connector in the CCU. Stryker has investigated and identified a service repair issue which contributed to the bent pins noted with these devices.
What to do
Contact the manufacturer if you require additional information.
The official notice
This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.
Read the official notice (opens in a new tab)