The recall affects the AGFA Digital Radiography X‑Ray system DR 800 with MUSICA Dynamic (common name DR 800), distributed nationwide in the U.S.; 35 units are involved. It was recalled because the Air Kerma rate (AKR) for pulsed fluoroscopy can exceed the maximum allowed 88 mGy/min if the frame rate is increased on the NX workstation while ABS (automatic brightness/beam stabilization) is not enabled.
What's being recalled
Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800
Recalling firm: AGFA Healthcare Corp., Greenville, SC, United States.
How to identify it
N/A
The hazard
It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
What to do