Skip to content

Health & beauty Canada Health product recall

AirSeal Access Port With Blunt Obturator and AirSeal Access Port With Optical Obturator

AirSeal Access Port with Blunt Obturator (models IASB12-100, IASB5-150, IASB12-120) and AirSeal Access Port with Optical Obturator (models IAS12-150LPI, IAS12-100LPI, IAS5-75LP, IAS8-100LP, IAS8-120LP, IAS5-100LP, IAS12-120LPI, IAS5-120LP) are affected; affected units are identified by those model/catalogue numbers and by lot or serial numbers (more than 10 numbers—contact the manufacturer). They were recalled because a prior manufacturing process update may cause erroneous overpressure alerts under certain conditions, potentially leading to loss of pneumoperitoneum and treatment delays; contact the manufacturer for additional information.

What's being recalled

Companies: Conmed Corporation, 525 French Rd., Utica, New York, United States, 13502-5994.

How to identify it

Affected products: AirSeal Access Port With Blunt Obturator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: IASB12-100

Affected products: AirSeal Access Port With Blunt Obturator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: IASB5-150

Affected products: AirSeal Access Port With Blunt Obturator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: IASB12-120

Affected products: AirSeal Access Port With Optical Obturator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: IAS12-150LPI

Affected products: AirSeal Access Port With Optical Obturator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: IAS12-100LPI

Affected products: AirSeal Access Port With Optical Obturator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: IAS5-75LP

Affected products: AirSeal Access Port With Optical Obturator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: IAS8-100LP

Affected products: AirSeal Access Port With Optical Obturator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: IAS8-120LP

Affected products: AirSeal Access Port With Optical Obturator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: IAS5-100LP

Affected products: AirSeal Access Port With Optical Obturator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: IAS12-120LPI

Affected products: AirSeal Access Port With Optical Obturator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: IAS5-120LP

The hazard

Complaint review identified a prior manufacturing process update that may cause erroneous overpressure alerts under certain conditions, potentially leading to loss of pneumoperitoneum and treatment delays. Image was added to IFU to reinforce proper device tightening. No injuries or deaths have been reported.

What to do

Contact the manufacturer if you require additional information.

The official notice

This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.

Read the official notice (opens in a new tab)

A little guidance