Skip to content

Health & beauty Canada Health product recall

Aisys, Avance, and Aespire Anesthesia Machines

What's being recalled

Companies: Datex-Ohmeda, Inc., 3030 Ohmeda Drive, Madison, Wisconsin, United States, 53707-7550.

How to identify it

Affected products: Aisys Carestation Anesthesia System - Main Unit; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalog number: 1011-9000-000

Affected products: Avance CS2 - Main Unit; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalog number: 1009-9050-000

Affected products: Aespire 7900 Smartvent Anesthesia System - Main Unit; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalog number: 1009-9012-000

Affected products: GE Datex-Ohmeda Avance Anesthesia System; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalog number: 1009-9002-000

Affected products: Aespire Anesthesia Machine; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalog number: 1009-9000-000

Affected products: Aisys CS2 Anesthesia Machine Main Unit; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalog number: 1011-9050-000

Affected products: Aespire View - Anesthesia Machine; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalog number: 1009-9212-000

The hazard

GE Healthcare has become aware of a potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

What to do

Contact the manufacturer if you require additional information.

The official notice

This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.

Read the official notice (opens in a new tab)

A little guidance

All guides

Search by name, description or tag. Press Enter to see every match.