Alaris & BD Alaris Point-Of-Care Unit (PCU)

Alaris and BD Alaris Point-of-Care Units (PCU), model 8015 and software v12.1.3 and prior (serial numbers with more than 10 digits—contact the manufacturer for specifics), are affected. Because labeling does not state which Transport Layer Security (TLS) versions must be enabled, the PCUs are only compatible with TLS 1.0; disabling TLS 1.0 can cause wireless network connectivity failures, interoperability failures requiring manual programming, and memory leaks that may lead to system error during use. Contact the manufacturer for additional information.

What's being recalled

Companies: Carefusion 303, Inc., 10020 Pacific Mesa Blvd., San Diego, California, United States, 92121.

How to identify it

Affected products: Alaris & BD Alaris Point-Of-Care Unit (PCU); Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 8015

The hazard

Becton Dickinson (BD) is issuing this voluntary product communication because its labeling does not indicate which Transport Layer Security (TLS) versions must be enabled on the hospital network to ensure secure wi-fi communication with the Alaris and BD Alaris™ Point-of-Care Unit (PCU) when using enterprise-level wi-fi security. Without clear labeling, hospitals may unknowingly disable TLS 1.0, resulting in network connectivity failure and memory leaks which could lead to potential system error.

Alaris™ and BD Alaris™ PCUs (v12.1.3 and prior) are not compatible with TLS security frameworks beyond TLS 1.0. If TLS 1.0 is disabled on the customer network, the PCUs will fail to connect to the wireless network. For customers utilizing bi-directional workflow, this leads to interoperability failures which require users to manually program devices.

If TLS 1.0 is disabled on the customer network during use, PCUs (V12.1.3 and prior) may encounter a memory leak when the authentication fails, which consumes over time, all available memory and leads to a system error. This may happen during active infusions.

Future versions of BD Alaris™ PCUs that are not yet available in Canada will be compatible with TLS 1.0, 1.1, and 1.2.

What to do

Contact the manufacturer if you require additional information.

The official notice

This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.

Read the official notice (opens in a new tab)

A little guidance

All guides

Search by name, description or tag. Press Enter to see every match.