Argon Medical Devices is recalling L‑Cath PICC S/L Peripherally Inserted Central Catheters (including 26ga REF 384539; 18ga REF 384464; Basic Kit 28ga REF 384516; 20ga REF 384463) and L‑Cath Midline S/L Catheter 20ga REF 384620 — identifiable by those REF/UDI‑DI/lot numbers and sizes. The catheters may leak or crack at the hub, which can require replacement or cause inability to flush/aspirate, catheter occlusion, therapy interruption, blood loss, tissue irritation, or bloodstream infection.
What's being recalled
L-Cath PICC S/L Peripherally Inserted Central Catheter: 26ga (1.9F) 0.60mm x 30cm, REF: 384539; 18ga (3.5F) 1.15mm x 60cm, REF: 384464; Basic Kit 28GA (1.2F) x 25cm 1 Lumen), REF: 384516; 20ga (3F) 1.00mm x 60cm, REF: 384463
L-Cath Midline S/L Catheter 20ga (3F) 1mm x 20cm, REF: 384620
Recalling firm: Argon Medical Devices, Inc, Athens, TX, United States.
How to identify it
REF/UDI-DI/Lot(Expiration): 384539/00886333209934/11488167(7/14/2026), 11488168(7/17/2026), 11494535E(8/1/2026), 11507766(11/13/2026), 11550780(12/12/2026), 11551487(12/15/2026), 11551488(12/15/2026); 384464/00886333209835/11490593(7/28/2026), 11494886(8/28/2026)11501150(10/6/2026); 384516/00886333209910/11500410(8/14/2026), 11501856(8/14/2026), 11504222(9/9/2026), 11506952(9/9/2026), 11523868(9/18/2026), 11551789(9/18/2026), 11553909(11/13/2026), 11559226(11/20/2026); 384463/00886333209828/11490797(8/1/2026), 11490959(8/11/2026) REF/UDI-DI/Lot(Expiration): 384620/00886333210008/11490543(7/28/2026)
The hazard
Catheters, indicated for short/long-term peripheral access to central venous system for the administration of fluids, medications, nutrients; and sampling, may experience leaking and/or cracking at the hub, which may lead to replacement, inability to flush/aspirate catheter, catheter occlusion, therapy interruption, blood loss, tissue irritation, bloodstream infection.
What to do
Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.
The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.
Where it was sold
US: AR, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OR, TN, TX, UT, VA, WA. OUS: MYS, CAN, SAU
The official notice
This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.
Read the official notice (opens in a new tab)