Health & beauty Canada Health product recall

Arjo Tenor Mobile Lift

Arjo Tenor Mobile Lift units (models KHA1000‑US, TEN.107, KHA1010‑US; affected units have lot or serial numbers described as “more than 10 numbers”—contact the manufacturer for specifics) are affected. A supplier‑related brake component defect in the lift actuator may cause loss of self‑locking and sudden uncontrolled descent under load, so all affected units must be removed from use immediately and remain out of service until the actuator is replaced by authorized Arjo service personnel. Contact the manufacturer for additional information.

What's being recalled

Companies: Arjohuntleigh Ab, Hans Michelsensgatan 10, Malmo, Sweden, 21120.

How to identify it

Affected products: Arjo Tenor Mobile Lift; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: KHA1000-US

Affected products: Arjo Tenor Mobile Lift; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: TEN.107

Affected products: Arjo Tenor Mobile Lift; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: KHA1010-US

The hazard

A supplier‑related brake component defect in the Tenor lift actuator may cause loss of self‑locking and sudden uncontrolled descent under load. All affected units must be removed from use immediately and remain out of service until actuator replacement is completed by authorized Arjo service personnel.

What to do

Contact the manufacturer if you require additional information.

The official notice

This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.

Read the official notice (opens in a new tab)