Auris Health’s MONARCH Platform bronchoscopy system (catalog numbers MON-000005-01, MON-000005-01R, MON-000006, MON-000006-RFB, MON-000008; software versions 2.2.2, 2.2.3, 2.2.4, 2.2.5, 4.1.1) is affected. The device was recalled because a software issue can reinitialize the application to the patient-left position after a restart, causing unanticipated robotic arm movement toward patient-left that may contact the patient or endotracheal tube.
What's being recalled
MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software
Recalling firm: Auris Health, Inc, Santa Clara, CA, United States.
How to identify it
Software versions: 2.2.2, 2.2.3, 2.2.4, 2.2.5, and 4.1.1. UDI-DI: 10810068810803, 10810068810988. Serial Numbers: 110024, 120088, 120057, 120046, 120058, 120015, 120160, 120068, 120081, 120056, 110028, 120128, 120017, 120089, 120075, 110016, 120031, 120169, 120113, 120168, 120055, 120014, 120060, 120174, 120195, 120026, 110038, 120124, 120194, 110027, 110035, 120123, 120091, 120007, 110022, 120006, 120152, 120004, 120120, 120041, 120106, 110034, 120126, 120082, 120151, 120051, 120020, 120153, 120193, 120170, 120108, 120037, 120074, 120199, 120130, 120095, 110026, 120096, 120009, 120111, 120022, 120054, 120163, 110045, 120027, 110039, 120090, 120141, 120072, 120155, 110049, 120143, 110047, 120125, 120052, 110052, 120100, 120050, 120098, 120045, 120131, 110020, 110030, 110032, 110021, 120013, 120105, 120136, 120192, 110017, 120029, 120139, 110031, 120137, 120053, 110029, 120028, 120034, 120085, 120019, 120049, 120209, 120144, 120161, 120001, 110033, 120145, 120127, 110046, 110043, 120011, 120070, 120076, 120134, 120117, 110025, 120077, 120149, 120191, 120102, 120002, 120156, 120010, 120171, 120003, 120023, 120146, 120188, 120043, 120071, 120042, 120196, 110051, 120065, 120172, 120116, 120150, 120066, 120166, 120033, 110013, 120079, 120008, 120086, 120032, 120073, 1806P1, 120197, 120080, 120078, 14, 120012, 110041, 120084, 120219, 110036, 120181, 120083, 120067, 120021, 120216, 120217, 120218, 120069, 120204, 120207, 120215, 120237, 120208, 120213, 120164, 120132, 120182
The hazard
Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.
What to do
Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.
The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.
Where it was sold
Worldwide - US Nationwide distribution in the states of MN, FL, NJ, NY, VA, WI, AZ, AL, MS, LA, MI, MA, MT, PA, WV, GA, DE, TX, CA, OH, IN, SC, CT, ND, MO, DC, IL, NE, ID, WA, NC, TN, OR, CO, SD, NM, VT, MD and the countries of Hong Kong, China, Canada.
The official notice
This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.
Read the official notice (opens in a new tab)