Brahms GmbH recalls B¿R¿A¿H¿M¿S KRYPTOR compact REACT

The Brahms KRYPTOR compact REACT (REF 89986) used with Brahms PlGF Plus KRYPTOR (REF 859075N) on KRYPTOR compact PLUS analyzers is recalled; affected lots are 40102+R78, 40102+R79 and 40102+R80 (PlGF REF 859075N: all lots). The recall was issued because customers observed quality control values out of range at QC Level 1 that, if quality controls are not performed according to the Instructions for Use, may be underestimated and could contribute to falsely elevated sFlt‑1/PlGF ratios and a higher‑risk classification for progression to preeclampsia with severe features.

What's being recalled

B¿R¿A¿H¿M¿S KRYPTOR compact REACT, REF: 89986, used with B¿R¿A¿H¿M¿S PlGF Plus KRYPTOR, REF: 859075N, and run on the B¿R¿A¿H¿M¿S KRYPTOR compact PLUS analyzers. in vitro diagnostic test for Placental Growth Factor

Recalling firm: Brahms GmbH, Hennigsdorf, Germany.

How to identify it

REF: 89986, UDI-DI: 04260157638976, Lots: 40102+R78, 40102+R79, 40102+R80. REF: 859075N, UDI-DI: 04260157632196, All Lots.

The hazard

Customers have observed quality control values out of range at the lowest quality control level (QC Level 1) for the PIGF assay which require troubleshooting steps to resolve. If Quality Controls are not performed according to the Instructions for Use, the underestimation of QC1 may go undetected and could contribute to falsely elevated sFlt-1/PlGF ratio results. A falsely elevated sFlt-1/PlGF ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.

What to do

Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.

The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.

Where it was sold

Worldwide distribution. US states: NC, TX, MN, UT, NY, FL, OH, IL, CT, CA, GA. Other countries: DE, AT, AL, AU, BG, BR, CA, CH, CZ, DK, EC, ET, ES, FR, GB, GR, HK, HU, ID, IL, IT, KW, KZ, LT, MK, ML, MX, NL, NO, PA, PL, PT, RO, RS, SE, SI, TH, TU, TW, UA, ZA