BS 3/8x3/32 L1.7 (Bubble Sensor for 3/8” X 3/32” tubing, length 1,7 m)

Maquet Cardiopulmonary GmbH BS 3/8x3/32 L1.7 bubble sensor (for 3/8” x 3/32” tubing, length 1.7 m; model/catalogue 70105.5720; lot/serial numbers listed as more than 10 numbers—contact manufacturer) is affected. Internal investigations found the connecting cable near the sensor can fail with excessive bending, causing poor contact and triggering temporary or permanent errors on connected medical devices. Contact the manufacturer for additional information.

What's being recalled

Companies: Maquet Cardiopulmonary Gmbh, Kehler Strasse 31, Rastatt, Germany, 76437.

How to identify it

Affected products: BS 3/8x3/32 L1.7 (Bubble Sensor for 3/8” X 3/32” tubing, length 1,7 m); Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 70105.5720

The hazard

Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Bubble Sensor. Excessive bending of the cable can lead to poor contact, which may trigger errors on the connected medical device. These errors can occur temporarily when the cable is moved or permanently if the connection is fully compromised.

What to do

Contact the manufacturer if you require additional information.