Cardinal Health Monoject Tuberculin Syringe, 1 mL Luer Lock (Catalog Number 1180100777; Lot Numbers 531164X and 531162X; UDI/DI 10884521013414, 20884521013411, 30884521013418) is affected. The outer packaging and blister pack labeling identify the product as a 1 mL Luer Lock tuberculin syringe, but the syringes inside are U-100 insulin syringes, prompting the recall.
What's being recalled
Cardinal Health Monoject Tuberculin Syringe, 1 mL Luer Lock, Catalog Number 1180100777
Recalling firm: Cardinal Health 200, LLC, Waukegan, IL, United States.
How to identify it
UDI/DI 10884521013414 (each), 20884521013411 (box), 30884521013418 (case), Lot Numbers: 531164X, 531162X
The hazard
The outer packaging and blister pack labeling identify the product as 1mL Luer Lock Tuberculin Syringe (Product Code 1180100777), however the syringes contained within the packages are U-100 Insulin syringes.
What to do
Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.
The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.
Where it was sold
Worldwide distribution - US Nationwide and the country of Canada.
The official notice
This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.
Read the official notice (opens in a new tab)