Beckman Coulter CellMek SPS (model C44603) with software version 2.3.90.0 is affected; identify units by SW Version: 2.3.90.0. It was recalled because a software error can cause samples to be dispensed unexpectedly close to the bottom of output tubes instead of being dispensed above the fluid surface. Contact the manufacturer if you require additional information.
What's being recalled
Companies: Beckman Coulter, Inc., 250 S. Kraemer Blvd., Brea, California, United States, 92821.
How to identify it
Affected products: CellMek SPS; Lot or serial number: SW Version: 2.3.90.0; Model or catalogue number: C44603
The hazard
Beckman Coulter has confirmed an error in SW version 2.3.90.0 that causes samples to be dispensed unexpectedly close to the bottom of the output tubes. The intended design is for samples to be dispensed above the fluid surface into the output tubes.
What to do
Contact the manufacturer if you require additional information.