Skip to content

Health & beauty Canada Health product recall

Checkcell®

Checkcell® (Immucor Inc.) lot 43257, model/catalogue numbers 0002225 and 0002224 is affected; some vials exhibited clot formation, hemolysis and unexpected darkening of the red blood cells. Investigation determined the clotting and abnormal darkening were associated with fungal growth within the vials. Temporary corrective actions include product containment activities and a product recall, so owners should follow the recall and containment instructions.

What's being recalled

Companies: Immucor Inc., 3130 Gateway Drive, Norcross, Georgia, United States, 30071.

How to identify it

Affected products: Checkcell®; Lot or serial number: 43257; Model or catalogue number: 0002225

Affected products: Checkcell®; Lot or serial number: 43257; Model or catalogue number: 0002224

The hazard

Some vials of Checkcell lot 43257 exhibited clot formation, hemolysis, and unexpected darkening of the red blood cells. Investigation determined that the clotting and abnormal darkening were associated with fungal growth within the vials.

What to do

Temporary corrective actions include product containment activities and a product recall.

The official notice

This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.

Read the official notice (opens in a new tab)

A little guidance

All guides

Search by name, description or tag. Press Enter to see every match.