Skip to content

Health & beauty Canada Health product recall

cobas® c 703 Analytical Unit

The cobas® c 703 Analytical Unit (model/catalogue 09502971001; all lots/serial numbers) is affected. Premature deterioration of the tubing in the reaction cell wash units can allow basic wash to drip into other reaction cells, potentially causing incorrect test results. Owners should contact Roche Diagnostics Service at laval.techinfo@roche.com.

What's being recalled

Companies: Roche Diagnostics GmbH, Sandhofer Strasse 116, Mannheim, Germany, 68305.

How to identify it

Affected products: cobas® c 703 Analytical Unit; Lot or serial number: All lots.; Model or catalogue number: 09502971001

The hazard

Premature deterioration to the tubing of the reaction cell wash units for the cobas® c 703 analytical unit can occur. In the event of damaged tubing, basic wash may drip into other reaction cells, potentially causing incorrect test results.

What to do

Roche Diagnostics Service: laval.techinfo@roche.com

The official notice

This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.

Read the official notice (opens in a new tab)

A little guidance

All guides

Search by name, description or tag. Press Enter to see every match.