The cobas® c 703 Analytical Unit (model/catalogue 09502971001; all lots/serial numbers) is affected. Premature deterioration of the tubing in the reaction cell wash units can allow basic wash to drip into other reaction cells, potentially causing incorrect test results. Owners should contact Roche Diagnostics Service at laval.techinfo@roche.com.
What's being recalled
Companies: Roche Diagnostics GmbH, Sandhofer Strasse 116, Mannheim, Germany, 68305.
How to identify it
Affected products: cobas® c 703 Analytical Unit; Lot or serial number: All lots.; Model or catalogue number: 09502971001
The hazard
Premature deterioration to the tubing of the reaction cell wash units for the cobas® c 703 analytical unit can occur. In the event of damaged tubing, basic wash may drip into other reaction cells, potentially causing incorrect test results.
What to do
Roche Diagnostics Service: laval.techinfo@roche.com
The official notice
This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.