What's being recalled
Companies: Integra Lifesciences Production Corporation, 11 Cabot Blvd., Mansfield, Massachusetts, United States, 02048.
How to identify it
Affected products: Codman® Disposable Perforator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 26-1221
Affected products: Codman® Disposable Perforator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 26-1222
Affected products: Codman® Disposable Perforator; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 26-1223
The hazard
On April 2025, Integra issued a voluntary recall for the 14mm Codman® Disposable Perforators due to an inadequate weld (proud weld) on specific lots of 14mm Codman® Disposable Perforators that can potentially cause the product to disassemble (break/separate).
To date, there have been no complaints or adverse events reported in relation to the 9mm and 11mm Codman® Disposable Perforators disassembly. Out of an abundance of caution, as the 9 and 11 mm perforators are manufactured using the same equipment and operating parameters that affected the 14mm Perforator, Integra has made the decision to expand the scope of the field action to include the 9mm and 11mm perforators.
What to do
Contact the manufacturer if you require additional information
The official notice
This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.
Read the official notice (opens in a new tab)