Accuray CyberKnife Treatment Delivery System units (model/catalogue 0660000) with serial numbers C0417, C0545, C0573, and C0351 are affected. During collimator housing drop-off or pickup over the Xchange table the system may interlock, stop motion, and prompt service contact. Remove the patient before troubleshooting, do not move the treatment robot using the in-room pendant, and contact Accuray service or the manufacturer for additional information.
What's being recalled
Companies: Accuray, Inc., 1209 Deming Way, Madison, Wisconsin, United States, 53717.
How to identify it
Affected products: CyberKnife® Treatment Delivery System; Lot or serial number: C0417; Model or catalogue number: 0660000
Affected products: CyberKnife® Treatment Delivery System; Lot or serial number: C0545; Model or catalogue number: 0660000
Affected products: CyberKnife® Treatment Delivery System; Lot or serial number: C0573; Model or catalogue number: 0660000
Affected products: CyberKnife® Treatment Delivery System; Lot or serial number: C0351; Model or catalogue number: 0660000
The hazard
During collimator housing drop-off or pickup over the Xchange® table, the system may interlock, stop motion, and prompt service contact. Remove the patient before troubleshooting, do not move the treatment robot using the in-room pendant, and contact accuray service.
What to do
Contact the manufacturer if you require additional information.