Digital Diagnostic Mobile X-Ray System GM85

The Digital Diagnostic Mobile X‑Ray System GM85 (model GM85, serial number S/N 5143M3UJ200001N) is affected. It was recalled because bolts connecting the tube head unit (THU) to the arm may loosen or break from metal fatigue, which could allow the THU to detach and pose a risk of serious injury; customers will be contacted to arrange an inspection and repair.

What's being recalled

Companies: Samsung Electronics Co., Ltd., 129, Samsung-Ro, Yeongtong-Gu, Suwon-Si, Gyeonggi-Do, South Korea, 16677.

How to identify it

Affected products: Digital Diagnostic Mobile X-Ray System GM85; Lot or serial number: S/N 5143M3UJ200001N; Model or catalogue number: GM85

The hazard

We have received a report regarding the loosening or breakage of bolts that connect the tube head unit (THU) to the arm. An investigation determined that this may be caused by metal fatigue resulting from accumulated stress over a long period of use. If all connecting bolts were to fail, it could potentially cause the THU to detach from the arm, posing a risk of serious injury to a patient or user.

What to do

Customers to be contacted to arrange visit for inspection and repair.