ForeSight™ Oximeter Cable (model HEMFSM10) by Edwards Lifesciences may have missing or mispositioned laterality identification labels on the green sensor cables, which can cause the channel 1 and channel 2 labels to be switched. That mislabeling can cause the left/right StO2 values displayed on the monitor to be reversed, potentially leading to incorrect diagnosis or delayed treatment. If labels are missing or the flag/sensor labels do not match the channel labels on the housing, contact Technical Support; if they match, no further action is required.
What's being recalled
Companies: Edwards Lifesciences LLC, 1 Edwards Way, Irvine, California, United States, 92614.
How to identify it
Affected products: ForeSight™ Oximeter Cable; Lot or serial number: Not applicable.; Model or catalogue number: HEMFSM10
The hazard
This recall is regarding inaccurate StO2 readings from the HEMFSM10 ForeSight™ Oximeter Cable due to mispositioning of the sensor cables. Edwards has identified that during manufacturing, the laterality identification labels may be missing or inaccurately placed on the device. There is potential that the labels on the green cables may have been switched, with the channel 1 cable labeled as channel 2, and the channel 2 cable labeled as channel 1. This will lead to a mismatch as to the left and right values displayed on the monitor and what the labels indicate on the cables. Should this potential issue occur during use, the values displayed on the monitor will be switched and will represent the opposite region being monitored. Therefore, there is a potential for incorrect diagnosis, leading to potential delay of appropriate treatment. If either of the labels are missing, or the flag label and/or the sensor label do not match the channel label on the housing, contact the Technical Support. If the flag and sensor labels correctly correspond to the channel labels, there is no further action required.
What to do
Contact the manufacturer if you require additional information.
The official notice
This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.
Read the official notice (opens in a new tab)