What's being recalled
Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
Recalling firm: Getinge Disinfection Ab, Vaxjo, N/A, Sweden.
How to identify it
Model Number: 88-5. UDI-DI Numbers: 07340153700109, 07340153700116. Software Version: GD 14080 Rev A. All Lot Numbers with electrical heating.
The hazard
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
What to do
Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.
The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.
Where it was sold
Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, Norway, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.
The official notice
This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.
Read the official notice (opens in a new tab)