HeartMate 3™ System Controller

Abbott Medical is recalling HeartMate 3™ System Controllers (including Low Flow 2.0 models and HeartMate 3 LVAS Implant Kit components) after an increase in customer complaints about the Backup Battery Fault Alarm, shown as a flashing yellow wrench on the controller user interface. Patients and caregivers should follow the patient handbook: if the Backup Battery Fault Alarm is active, ensure no other alarms are active, then call their hospital contact for diagnosis and instructions and do not attempt to exchange the controller without clinical guidance; contact the manufacturer for additional information.

What's being recalled

Companies: Abbott Medical, 6035 Stoneridge Drive, Pleasanton, California, United States, 94588.

How to identify it

Affected products: HeartMate 3™ System Controller; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 106531US

Affected products: HeartMate 3™ System Controller Low Flow 2.0; Lot or serial number: s/n: HSC-138822; Model or catalogue number: 106531LF2

Affected products: HeartMate 3™ System Controller Low Flow 2.0; Lot or serial number: s/n: HSC-122360; Model or catalogue number: 106531LF2

Affected products: HeartMate 3™ LVAS Implant Kit; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 106524US

The hazard

Abbott is initiating a voluntary medical device recall for the HeartMate II® and HeartMate 3™ System Controllers. This recall does not involve device removal from where they are used or sold.

Abbott has observed an increase in customer complaints about the controller Backup Battery Fault Alarm. This alarm is indicated by a flashing yellow wrench on the user interface of the System Controllers. When the controller is connected to an external power source (e.g. 14V batteries or AC power via the Mobile Power Unit), this is an advisory alarm and does not indicate a critical system failure or impact pump functionality.

To reduce future instances of the Backup Battery Fault Alarm, Abbott is advising that when performing backup battery installation or replacement, clinicians should minimize excessive handling, movement, stress and pulling at the interface between the ribbon cable and controller. An example of excessive handling would be using the ribbon cable to hold the weight of the controller.

Please remind patients and caregivers to follow the patient handbook: if System Controller Backup Battery Fault Alarm is active, patients should first ensure that no other alarms are active, and then they should call their hospital contact as soon as possible for diagnosis and instructions. Patients and caregivers should not immediately attempt to exchange the system controller without clinical guidance.

What to do

Contact the manufacturer if you require additional information.

The official notice

This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.

Read the official notice (opens in a new tab)