HeartMate II LVAS System Controllers (model 106762; s/n PCX-22221, PCX-22321, PCX-22270, PCX-22271) are affected and can show a flashing yellow wrench Backup Battery Fault Alarm on the user interface when connected to external power. Abbott is recalling these controllers because of increased customer complaints about that advisory alarm. Patients should follow the patient handbook: if the alarm is active, ensure no other alarms are present and call their hospital contact as soon as possible; contact the manufacturer for more information.
What's being recalled
Companies: Thoratec Corporation, 6035 Stoneridge Drive, Pleasanton, California, United States, 94588.
How to identify it
Affected products: HeartMate II® LVAS - System Controller; Lot or serial number: s/n: PCX-22221; Model or catalogue number: 106762
Affected products: HeartMate II® LVAS - System Controller; Lot or serial number: s/n: PCX-22321; Model or catalogue number: 106762
Affected products: HeartMate II® LVAS - System Controller; Lot or serial number: s/n: PCX-22270; Model or catalogue number: 106762
Affected products: HeartMate II® LVAS - System Controller; Lot or serial number: s/n: PCX-22271; Model or catalogue number: 106762
The hazard
Abbott is initiating a voluntary medical device recall for the HeartMate II® and HeartMate 3™ System Controllers. This recall does not involve device removal from where they are used or sold.
Abbott has observed an increase in customer complaints about the controller Backup Battery Fault Alarm. This alarm is indicated by a flashing yellow wrench on the user interface of the System Controllers. When the controller is connected to an external power source (e.g. 14V batteries or AC power via the Mobile Power Unit), this is an advisory alarm and does not indicate a critical system failure or impact pump functionality.
To reduce future instances of the Backup Battery Fault Alarm, Abbott is advising that when performing backup battery installation or replacement, clinicians should minimize excessive handling, movement, stress and pulling at the interface between the ribbon cable and controller. An example of excessive handling would be using the ribbon cable to hold the weight of the controller.
Please remind patients and caregivers to follow the patient handbook: if System Controller Backup Battery Fault Alarm is active, patients should first ensure that no other alarms are active, and then they should call their hospital contact as soon as possible for diagnosis and instructions. Patients and caregivers should not immediately attempt to exchange the system controller without clinical guidance.
What to do
Contact the manufacturer if you require additional information.