Insulet Corporation recalls Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2.… and 2 more products

Omnipod 5 Pods, Omnipod DASH Pods, and Omnipod (Eros) Pods with the listed Model/REF numbers (for example POD-OMNI-I1-6720, POD-OMNI-I1-6220, POD-BLE-I1-520, POD-ZXP420) and associated lot/UDI identifiers are affected and can be recognised by those model/REF, lot and UDI details. They were recalled because external soft cannula damage during manufacturing can cause insulin to leak from the Pod instead of being delivered, leading to pump under-delivery and potentially high blood glucose, diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or death.

What's being recalled

Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). 4. OMNI-I1-6220. 5. OMNI-I1-6720. 6. BLE-H1-529 (10-Pack), BLE-H1-520 (Single Pod).

Omnipod DASH Pod. Includes the below Model/REF Numbers: 1. POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). 2. POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). 3. BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). 4. BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod).

Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). 2. ZXR425 (10-Pack), ZXR420 (Single Pod). 3. ZXP425 (10 Pack), ZXP420 (Single Pod).

Recalling firm: Insulet Corporation, Acton, MA, United States.

How to identify it

1. Model/REF Number: POD-OMNI-I1-6720. Pack UDI-DI: 20385083000579. Single Pod UDI-DI: 10385083000527. Lot Numbers: PH1U01032521, PH1U01102521, PH1U12102421, PH1U12172421, PH1U11212421, PH1U11252421, PH1U11222421, PH1U12092421, PH1U09112421, PH1U11262421, PH1U04032521, PH1U12022421, PH1U12182421, PH1U04302524, PH1U05022521, PH1U01222521, PR1U11202533, PH1U09122421, PH1U10102421, PH1U12122421, PH1U01302522, PH1U02222521, PH1U05052521, PH1U10042421, PH1U12102412, PH1U01312521, PH1U03312521, PH1U05012521, PH1U09122422, PH1U09162421, PH1U09192421, PH1U09212421, PH1U09262411, PH1U09262421, PH1U10112422, PH1U10122421, PH1U10172421, PH1U12132422, PH1U01132521, PH1U01172521, PH1U02132511, PH1U03202521, PH1U03262521, PH1U04042521, PR1U11192531, PH1U11192511, PR1U01162621, PH1U09172421, PH1U09182421, PH1U09202421, PH1U10142421, PH1U10152421, PH1U10162421, PH1U01232522, PH1U01312511, PH1U02032521, PH1U02042521, PH1U02212521, PH1U03282521, PH1U04092521, PH1U04102521, PH1U04112521, PH1U10172522, PH1U10182521, PH1U10202521, PH1U10212521, PH1U10222521, PH1U10232521, PH1U10242521, PH1U10252521, PH1U11152521, PH1U11172511, PH1U11172521, PH1U11182521, PR1U11182531, PR1U11182541, PR1U11192541, PH1U11192521, PR1U11192532, PR1U11192542, PR1U11202521, PR1U11202532, PR1U11202541, PR1U11202542, PR1U11212521, PR1U11222531, PR1U11222541, PR1U11222521, PR1U11232541, PR1U11242521, PR1U11242531, PR1U11252541, PR1U11252512, PR1U11252521, PR1U11262521, PR1U12012521, PR1U12022521, PR1U12032521, PR1U12102522, PR1U12112521, PR1U12122521, PR1U12132511, PR1U12132521, PR1U12152521, PR1U12162521, PR1U12172521, PR1U01132621, PR1U01152621, PR1U01172621, PR1U01202621, PR1U01212621, PR1U01222621, PR1U01232621, PR1U01272621, PR1U01302641, PR1U01312631, PR1U02022641, PR1U02032631, PR1U02032641, PR1U02042641, PR1U02052631, PR1U02052641, PR1U02062631, PR1U02112641, PR1U02132611, PR1U02132641, PR1U11252531, PR1U11262541, PH1U06282422. 2. Model/REF Number: POD-OMNI-I1-6220. Pack UDI-DI: 20385083000425. Single…

The hazard

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

What to do

Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.

The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.

Where it was sold

Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.

The official notice

This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.