Kiwi® Vacuum Assisted Delivery System

The Kiwi® Vacuum Assisted Delivery System (model VAC-6000MT; for lot or serial number information contact the manufacturer) is affected. The device may break at the Traction Force Indicator (TFI) joint during use when traction is applied at approximately a 45-degree angle. The manufacturer will contact customers with instructions on return of affected products.

What's being recalled

Companies: Clinical Innovations, LLC, 747 West 4170 South, Murray, Utah, United States, 84123.

How to identify it

Affected products: Kiwi® Vacuum Assisted Delivery System; Lot or serial number: Contact the manufacturer.; Model or catalogue number: VAC-6000MT

The hazard

The device may break at the Traction Force Indicator (TFI) joint during use when traction is applied at approximately a 45-degree angle.

What to do

Manufacturer will contact custmers with instructions on return of affected products.

The official notice

This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.