Likorall patient lift systems from Liko AB are affected — affected models include Likorall 200 (cat. no. 3121001), 250 S (3124050), multiple 242 S and 242 ES R2R variants (e.g. 3122009, 3122010, 3122007, 3122008 and NA variants), 250 ES variants (e.g. 3122501, 3122502 and NA variants) and 243 ES variants (e.g. 3123001, 3123002); affected units have lot/serial numbers with more than 10 numbers, contact the manufacturer to confirm. The recall is due to a potential motor shaft breakage combined with reduced single fault safety system performance that in rare double-fault cases has allowed the lift strap to lower more quickly than expected. Contact the manufacturer if you require additional information.
What's being recalled
Companies: Liko AB, Nedre Vagen 100, Lulea, Sweden, 97592.
How to identify it
Affected products: Likorall™ 200; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3121001
Affected products: Likorall™ 250 S; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3124050
Affected products: Likorall™ 242 S; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122009
Affected products: Likorall™ 242 S; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122010
Affected products: Likorall™ 242 S; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122009NA
Affected products: Likorall™ 242 S; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122010NA
Affected products: Likorall™ 250 ES; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122502
Affected products: Likorall™ 250 ES; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122502NA
Affected products: Likorall™ 250 ES; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122501
Affected products: Likorall™ 250 ES; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122501NA
Affected products: Likorall™ 242 ES R2R; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122008NA
Affected products: Likorall™ 242 ES R2R; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122007
Affected products: Likorall™ 242 ES R2R; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122007NA
Affected products: Likorall™ 242 ES R2R; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122008
Affected products: Likorall™ 243 ES; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3123001
Affected products: Likorall™ 243 ES; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3123002
Affected products: Likorall™ 243 ES; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3123001NA
Affected products: Likorall™ 243 ES; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3123002NA
Affected products: Likorall™ 242 S R2R; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122012
Affected products: Likorall™ 242 S R2R; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122012NA
Affected products: Likorall™ 242 S R2R; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122011
Affected products: Likorall™ 242 S R2R; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122011NA
Affected products: Likorall™ 242 ES; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122006
Affected products: Likorall™ 242 ES; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122005
Affected products: Likorall™ 242 ES; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122006NA
Affected products: Likorall™ 242 ES; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 3122005NA
The hazard
Baxter Corporation is issuing an urgent medical device correction for the Likorall™ ceiling-mounted overhead patient lift system due to a potential issue involving motor shaft breakage combined with reduced performance of the single fault safety (SFS) system. In rare cases where a double fault failure occurs simultaneously, with both a motor shaft breakage and reduced SFS system performance, customers have reported instances of the lift strap lowering more quickly than expected during use.
What to do
Contact the manufacturer if you require additional information.
The official notice
This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.
Read the official notice (opens in a new tab)