Medline Industries, LP recalls Airway Exam Kit and 6 more products

Multiple Medline convenience and procedure kits are affected, including the Airway Exam Kit (DYKE1796) and numerous bronchoscopy, BAL, endoscopy, thoracic and hip kits identified by listed part numbers and lot/UDI codes. They were recalled because the kits contain Olympus single‑use biopsy valves that may shed rubber fragments from the valve slit, which could become a foreign body in the tracheobronchial tree and potentially cause inflammatory response, hypoxia, prolongation of the procedure or require intervention for removal.

What's being recalled

Airway Exam Kit, DYKE1796

Medline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2. NON MB BRONCHOPSCOPY CCH LH PK, DYNJT6182 3. PULMONARY BRONCH PACK, DYNDA2861A

THORACIC ROBOTS, DYNJ908777B

Medline kits containing Olympus biopsy valves: 1. BAL KIT, DYNDA2137B 2. ENDO KIT, DYKE1915

Medline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY, DYNJ900898I DYNJ901922G 2. FLEXIBLE BRONCHOSCOPY KIT, DYKM1417I DYKM1417J 3. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146A 4. RIGID KIT, DYNDA2138A

Medline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750I 5. KIT, WR THORACIC, DYNJ911252 6. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146B 7. LEX THORACIC, DYNJ902016K DYNJ902016L 8. ORGAN PROCUREMENT, DYNJ908686 9. PULMONARY PACK, DYKE1859 10. THORACIC, DYNJ901666L 11. THORACIC ROBOTICS, DYNJ908777A 12. THORACOTOMY/THORACOSCOPY, DYNJ900482N DYNJ900482O 13. XPS, DYNJ907605

MEDLINE ANTERIOR HIP PACK DYNJ64672B

Recalling firm: Medline Industries, LP, Northfield, IL, United States.

How to identify it

UDI-DI 10193489874716 (ea) 40193489874717 (case) Lots 24CBT023 24FBB701 24LBL343 25GBP141 1. UDI-DI 10198459629334 (EA) 40198459629335 (CS) LOTS 26AMG895 2. UDI-DI 10198459545283 (EA) 40198459545284 (CS) LOTS 25LBJ366 26ABR365 3. UDI-DI 10195327471859 (EA) 40195327471850 (CS) LOTS 24EBA317 24EBV431 24IBD694 24JBQ783 UDI-DI 10198459269196 (EA) 40198459269197 (CS) LOTS 25EBD732 25GBG733 25IBQ613 25JBD457 25KBT352 26ABA235 26ABP735 1. UDI-DI 10195327501259 (EA) 40195327501250 (CS) LOTS 24DBW065 24EBV301 24HBB492 24JBW143 25ABG413 25CBH359 25DBD213 25GBP569 25HBQ088 25KBF725 26ABE643 2. UDI-DI 10195327371029 (EA) 40195327371020 (CS) LOTS 24FBM018 24GBL319 1. DYNJ900898I UDI-DI 10198459273513 (EA) 40198459273514 (CS) LOTS 25DME064 25HMD841 25JMD489 25KME329 25KMH848 25LMC029 25LMH235 25LMK305 DYNJ901922G UDI-DI 10195327640422 (EA) 40195327640423 (CS) LOTS 24KBP812 2. DYKM1417I UDI-DI 10195327456757 (EA) 40195327456758 (CS) LOTS 24DLA405 24ELA912 24GLA871 24ILA151 24JLA326 DYKM1417J UDI-DI 10198459191640 (EA) 40198459191641 (CS) LOTS 25BLA761 25BLA886 25DLA353 25ELA056 25FLA429 25HLA196 25HLA242 25KLA388 25KLA610 3. DYNJ58146A UDI-DI 10193489611977 (EA) 40193489611978 (CS) LOTS 23KDA851 24ADC096 24CDB107 24DDB887 24IDA855 24JDB494 24LDA691 25BMF825 25DMD969 25FMC760 25GMD422 4. DYNDA2138A UDI-DI 10193489874594 (EA) 40193489874595 (CS) 24DBM325 24DBM383 25DBK159 1. DYKE2091 UDI-DI 10198459606397 (EA) 40198459606398 (CS) LOT 25LBR616 2. DYNJ900898G UDI-DI 10195327479473 (EA) 40195327479474 (CS) LOTS 23JBI824 23KBJ535 24ABR789 24ABV917 24BBU610 24DMK530 24FMF039 24GME122 24JMB670 24KMH657 25AMA403 25BME938 DYNJ901922I UDI-DI 10198459237270 (EA) 40198459237271 (CS) LOTS 25BBL628 25FBM655 25HBW072 25KBO230 3. DYNJ38313B UDI-DI 10884389857700 (EA) 40884389857701 (CS) LOTS 24BBR676 24CBF032 24FBD924 24JBE567 25EBA277 25FBO580 25IBI418 4. DYNJ32750G UDI-DI 10889942212491 (EA) 40889942212492 (CS) LOTS 23KME954 23LMD492 24BMF722 24FMH267 24HMC509 DYNJ32750I UDI-DI 10198459245695…

The hazard

Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

What to do

Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.

The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.

Where it was sold

US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.

The official notice

This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.

Read the official notice (opens in a new tab)

A little guidance

All guides

Search by name, description or tag. Press Enter to see every match.