Midmark IQecg Digital ECG (models/catalog numbers 4-000-0061 and 4-000-0062) and Midmark Digital ECG HD+12 and HD+15 (model 4-000-0080) running IQecg plugin software versions starting with v10 and prior to v10.1.0.18 are affected; the recall also identifies specific serial numbers and UDI-DIs listed in the notice. They were recalled because a communication anomaly may result in incorrect electrocardiogram waveforms and an inaccurate beats-per-minute calculation.
What's being recalled
Midmark IQecg Digital ECG, Model/Catalog Number: 4-000-0061, 4-000-0062 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.
Midmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.
Recalling firm: Midmark Corporation, Versailles, OH, United States.
How to identify it
Lot Code: This correction affects IQecg plugin software versions starting with v10 and prior to v10.1.0.18 installed on Midmark IQecg Digital ECG. The Midmark IQecg Digital ECG system information is provided below. The total quantities manufactured and distributed noted below are the combined IQecgs Digital ECG and Midmark Digital ECG systems. Serial Numbers: 480000 4820704 UDI-DI: 00841709100581, 00841709100321, 00841709100895, 00841709100901 Lot Code: This correction affects IQecg plugin software versions starting with v10 and prior to v10.1.0.18 installed on Midmark Digital ECG. The Midmark Digital ECG system information is provided below. The total quantities manufactured and distributed noted below are the combined IQecgs Digital ECG and Midmark Digital ECG systems. Serial Numbers: 7817519 7817645, L1130587 V2967564 UDI-DI: 00841709121319, 08056732652491
The hazard
A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.
What to do
Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.
The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.
Where it was sold
Worldwide distribution - US Nationwide and the country of Canada.
The official notice
This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.
Read the official notice (opens in a new tab)