Olympus Corporation of the Americas recalls Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use… and 1 more products

Olympus Single Use Electrosurgical Snare SD-400 devices (models SD-400U-10, UDI 04953170408243, and SD-400U-15, UDI 04953170408250) used with endoscopes for removal/cauterization of gastrointestinal polyps are affected; lots listed include 27V–57V and variants (see notice for full lot list). They were recalled because sterility may be compromised by small holes occurring during packaging into the shelf carton or transportation. (No patient/user actions were specified in the notice.)

What's being recalled

Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical Snare Model/Catalog Number: SD-400U-10 Product Description: These instruments have been designed to be used with an Olympus endoscope for the removal and/or cauterization of diminutive polyps, sessile polyps, pedunculated polyps, and tissue from within the GI tract.

Brand Name: Olympus Single Use Single Use Electrosurgical Snare SD-400 Product Name: Single Use Single Use Electrosurgical Snare Model/Catalog Number: SD-U400U-15 Product Description: These instruments have been designed to be used with an Olympus endoscope for the removal and/or cauterization of diminutive polyps, sessile polyps, pedunculated polyps, and tissue from within the GI tract.

Recalling firm: Olympus Corporation of the Americas, Center Valley, PA, United States.

How to identify it

Model: SD-400U-10; Lot Numbers: 27V, 28V, 29V, 2XV, 2YV, 2ZV, 31V, 32V, 33V, 34V, 35V, 36V, 37V, 38V, 39V, 3XV, 3YV, 3ZV, 41V, 42V, 43V, 44V, 45V, 46V, 47V, 48V, 49V, 4XV, 4YV, 51V, 52V, 53V, 54V, 55V, 56V, 57V UDI: 04953170408243 Model: SD-400U-15; Lot Numbers: 27V, 28V, 29V, 2XV, 2YV, 2ZV, 31V, 32V, 33V, 34V, 35V, 36V, 37V, 38V, 39V, 3XV, 3YV, 3ZV, 41V, 42V, 43V, 44V, 45V, 46V, 47V, 48V, 49V, 4XV, 4YV, 4ZV, 51V, 52V, 53V, 54V, 55V, 56V UDI: 04953170408250

The hazard

Sterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton.

What to do

Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.

The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.

Where it was sold

Worldwide distribution - US Nationwide and the countries of GU, and MP.