What's being recalled
Philips CT systems labeled as the following: 1. Brilliance CT 40 Channel, Model Number: 728235. 2. Brilliance CT 64 Channel, Model Number: 728231. 3. Ingenuity Flex, Model Number: 728317
Philips CT systems labeled as the following: 1. Brilliance 16 Power, Model Number: 728240. 2. Brilliance CT 16 Slice, Model Number: 728246. 3. Brilliance CT 6 Slice, Model Number: 728256.
Philips CT systems labeled as the following: 1. Big Bore RT, Model Number: 728242. 2. Brilliance CT Big Bore Oncology, Model Number: 728243. 3. Brilliance CT Big Bore, Model Number: 728244.
Philips CT systems labeled as the following: 1. Brilliance iCT, Model Number: 728306.
Philips CT systems labeled as the following: 1. Brilliance iCT SP, Model Number: 728311.
Philips CT systems labeled as the following: 1. Ingenuity CT 728326, Model Number: 728321. 2. Ingenuity CT 728326, Model Number: 728323. 3. Ingenuity CT, Model Number: 728326.
Philips CT systems labeled as the following: 1. IQon Spectral CT, Model Number: 728332.
Philips CT systems labeled as the following: 1. Ingenuity Core128 / Elite China, Model Number: 728324.
Philips CT systems labeled as the following: 1. Ingenuity CT Family Brazil, Model Number: 728325.
Philips CT systems labeled as the following: 1. CT 6000, Model Number: 728307.
Recalling firm: Philips North America Llc, Cambridge, MA, United States.
How to identify it
1. Model Number: 728235; UDI-DI: N/A; Serial Numbers: 9223, 9115, 9225, 29144. 2. Model Number: 728231; UDI-DI: N/A; Serial Numbers: 95647, 9194, 95341, 95234, 10051, 95488, 10098, 95695, 10309, 95217, 95698, 10443, 10544, 10531, 10712, 10662, 95834, 9990, 9784, 95632, 9058, 9089, 9522, 9580, 9688, 9877, 9806, 9808, 9671, 9694, 9703, 9830, 29068, 9821, 9869, 9837, 90144, 95203, 95225, 95303, 95328, 95362, 95402, 95395, 9852, 95492, 95491, 9965, 9974, 95453, 95502, 10006, 10025, 95520, 10060, 95630, 95621, 10141, 95673, 10153, 10137, 95521, 90128, 9605, 10218, 10199, 10247, 10275, 10267, 10245, 10394, 10307, 90124, 10281, 95692, 29045, 95349, 95639, 10344, 10330, 10310, 10348, 10391, 10393, 10418, 10205, 9799, 10429, 29062, 10511, 10431, 10515, 10535, 10315, 10521, 10518, 10566, 10536, 10522, 10496, 10560, 10568, 90170, 10433, 10574, 10609, 10559, 10468, 95725, 10622, 95710, 95729, 10590, 10628, 10683, 10668, 10666, 10607, 95758, 10706, 95776, 95821, 95757, 95858, 95795, 9868, 10729, 95791, 10734, 10718, 95807, 10649, 95781, 95790, 95823, 95813, 10788, 10774, 95759, 95852, 95848, 95116, 95872, 95871, 95914, 95886, 95498, 95054, 95928, 95933, 95025, 90162, 95144, 29039, 95932, 95399, 10031, 90804, 10161, 10494, 9934, 90161. 3. Model Number: 728317; UDI-DI: N/A; Serial Numbers: 345040, 345053, 345155, 345031, 345048, 345056, 345055, 345063, 345050, 345100, 345117, 345122, 345133, 345131, 345171. 1. Model Number: 728240; UDI-DI: N/A; Serial Number: 6022. 2. Model Number: 728246; UDI-DI: N/A; Serial Numbers: 5029, 5698, 5834, 6272, 6313, 5307, 5305, 50129, 5286, 50017, 5388, 5424, 5422, 5515, 5497, 5584, 5668, 5624, 5771, 5814, 5872, 5907, 5900, 5928, 5869, 5939, 5944, 5993, 6137, 5411, 6160, 6153, 5504, 6182, 6199, 6209, 6219, 6271, 6302, 6330, 6357, 6358, 6369, 6405, 6343, 6410, 6558, 6490, 6635, 6532, 6590, 6547, 6669, 6691, 6563, 6756, 6755, 6738, 3578, 6776, 260046, 260074, 6752, 50219, 50028, 5412, 5839. 3. Model Number: 728256; UDI-DI: N/A; Serial Numbers…
The hazard
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
What to do
Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.
The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.
Where it was sold
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauritius, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Vietnam.
The official notice
This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.