Philips Respironics Trilogy Evo ventilator devices (Trilogy Evo, Trilogy Evo O2, Trilogy EV300, Trilogy Evo Universal) running Software Version 1.05.15.00 and identified by all model numbers, UDIs and serial numbers are affected. They were recalled because using non‑pneumatic nebulizers with Trilogy Evo platform ventilators may cause a discrepancy between the set tidal volume and the tidal volume received by the patient.
What's being recalled
Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.
Philips Respironics Trilogy Evo O2, Software Version 1.05.15.00. Continuous home-use ventilator device.
Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.
Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.
Recalling firm: Philips Respironics, Inc., Murrysville, PA, United States.
How to identify it
All Model No.; All UDI; All Serial No.
The hazard
Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
What to do
Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.
The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.
Where it was sold
Worldwide distribution: US (Nationwide) and OUS (Foreign) to countries of: Argentina, Austria, Australia, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bolivia, Brazil, Bahamas, Canada, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, France, Georgia, French Guiana, Gibraltar, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jersey, Jordan, Japan, Kenya, South Korea, Kuwait, Lebanon, Lithuania, Morocco, Montenegro, North Macedonia, Myanmar, Martinique, Malta, Mexico, Malaysia, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, French Polynesia, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, R¿union, Romania, Serbia, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Thailand, Tunisia, Turkey, Taiwan, Ukraine, United Arab Emirates, United Kingdom, Vietnam, Mayotte, South Africa, Zimbabwe.
The official notice
This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.
Read the official notice (opens in a new tab)