What's being recalled
Biograph mMR. Model Number: 10433372.
BIOGRAPH One (DE). Model Number: 11689172.
MAGNETOM Cima.X (DE). Model Number: 11647158.
MAGNETOM Connectom.X. Model Number: 11371480.
MAGNETOM Lumina (DE). Model Number: 11344916.
MAGNETOM Prisma. Model Number: 10849582.
MAGNETOM Skyra (DE). Model Number: 10432915. 510(k) Numbers: K101347, K111242, K123510, K133435, K140253, K141977, K142515, K153343, K173592, K202014, K231560.
MAGNETOM Skyra fit. Model Number: 10849580.
MAGNETOM Skyra Fit BioMatrix. Model Number: 11516217.
MAGNETOM Spectra as the following: 1. MAGNETOM Spectra. Model Number: 10837643. 2. MAGNETOM Spectra (CN). Model Number: 10655588.
MAGNETOM Verio. Model Number: 10276755.
MAGNETOM Verio Dot. Model Number: 10684333.
MAGNETOM Verio Dot Upgrade. Model Number: 10684334.
MAGNETOM Vida (DE). Model Number: 11060815. 510(k): K170396, K173617, K181433, K183254, K192924, K203443, K213693, K213805, K220425, K231560.
MAGNETOM Vida Fit. Model Number: 11410481.
Recalling firm: Siemens Medical Solutions USA, Inc, Malvern, PA, United States.
How to identify it
Model Number: 10433372. UDI Number: (01)04056869006741(21)151032, (01)04056869006741(21)151057, (01)04056869006741(21)510535, (01)04056869006741(21)151034, (01)04056869006741(21)51003, (01)04056869006741(21)151021, (01)04056869006741(21)51055, (01)04056869006741(21)151013, (01)04056869006741(21)51064, (01)04056869006741(21)151044, (01)04056869006741(21)51038, (01)04056869245188(21)51006, (01)04056869006741(21)51030, (01)04056869006741(21)151004, (01)04056869006741(21)51022, (01)04056869006741(21)151049, (01)04056869006741(21)51007, (01)04056869006741(21)151045, (01)04056869006741(21)51032, (01)04056869006741(21)51057, (01)04056869006741(21)51015, (01)04056869006741(21)151025, (01)04056869006741(21)51048, (01)04056869006741(21)51062, (01)04056869006741(21)51010, (01)04056869006741(21)51051. Serial Numbers: 151032, 151057, 51053, 151034, 51003, 151021, 51055, 151013, 51064, 151044, 51038, 51006, 51030, 151004, 51022, 151049, 51007, 151045, 51032, 51057, 51015, 151025, 51048, 51062, 51010, 51051. Additional serial number added 01/23/2026: 51007. Model Number: 11689172. UDI Numbers: (01)4068151020472(21)238008. Serial Number: 238008. Model Number: 11647158. UDI Numbers: (01)04056869975221(21)213029, (01)04056869975221(21)213006, (01)04056869975221(21)213062, (01)04056869975221(21)213043, (01)04056869975221(21)213041, (01)04056869975221(21)213034, (01)04056869975221(21)213035, (01)04056869975221(21)213028, (01)04056869975221(21)213031, (01)04056869975221(21)213018, (01)04056869975221(21)213061, (01)04056869975221(21)213005, (01)04056869975221(21)213021, (01)04056869975221(21)213022, (01)04056869975221(21)213056, (01)04056869975221(21)213030, (01)04056869975221(21)213039, (01)04056869975221(21)213025, (01)04056869975221(21)213063, (01)04056869975221(21)213057, (01)04056869975221(21)213026, (01)04056869975221(21)213020. Serial Numbers: 213029, 213006, 213062, 213043, 213041, 213034, 213035, 213028, 213031, 213018, 213061, 213005, 213021, 213022, 213056, 213030, 213039…
The hazard
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
What to do
Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.
The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.
Where it was sold
Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyprus, Czechia, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Ethiopia, Finland, Faroe Islands, France, United Kingdom of Great Britian and Northern Ireland, Georgia, Greece, Hong Kong, Croatia, Hungary, Canary Islands, Indonesia, Ireland, Israel, India, Iran, Iceland, Italy, Jordan, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Lithuania, Luxembourg, Latvia, Morocco, Monaco, Moldova, Montenegro, Republic of North Macedonia, Myanmar, Mongolia, Macao, Malta, Maldives, Mexico, Malaysia, Mozambique, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Senegal, Syrian Arab Republic, Thailand, Turkmenistan, Tunisia, Turkey, Trinidad and Tobago, Taiwan, Tanzania, Ukraine, Uganda, Uzbekistan, Viet Nam, Kosovo, South Africa.
The official notice
This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.