Skip to content

Health & beauty Canada Health product recall

SLSolution™

SLSolution™ reagent (Copan Italia S.P.A.), models/catalogue numbers 0U019N (lots 2415124, 2504562, 2500451, 2502718) and 0U020N.A (contact the manufacturer for affected lot numbers) is affected. The reagent must be colourless; it was recalled because an unexpectedly high number of devices showed pink or yellow discoloration from oxidized dithiothreitol (DTT), which may impair fluidification of respiratory specimens. Contact the manufacturer for additional information.

What's being recalled

Companies: Copan Italia S.P.A., Via F. Perotti, 10, Brescia, Italy, 25125.

How to identify it

Affected products: SLSolution™; Lot or serial number: 2415124; Model or catalogue number: 0U019N

Affected products: SLSolution™; Lot or serial number: 2504562; Model or catalogue number: 0U019N

Affected products: SLSolution™; Lot or serial number: 2500451; Model or catalogue number: 0U019N

Affected products: SLSolution™; Lot or serial number: 2502718; Model or catalogue number: 0U019N

Affected products: SLSolution™; Lot or serial number: Contact the manufacturer.; Model or catalogue number: 0U020N.A

The hazard

SLSolution™ reagent includes dithiothreitol (DTT), which is oxidized when it gets in contact with air and/or metals, potentially losing its functionality in fluidification of respiratory specimens.

SLSolution™ reagent must show a colourless appearance. On the other hand, when the DTT reagent is deteriorated:

the SLSolution™ reagent shows a pink or yellow discoloration (device failure),

its functionality in respiratory specimens fluidification cannot be guaranteed.

For this reason:

SLSolution™ devices are manufactured to guarantee the stability and activity of DTT,

a dedicated warning is included in devices' instructions for use (ref. EIFU013, warning number 8) to instruct the user to discard any non-colourless piece ("do not use Copan SLSolution™ if the reagent is not clear/colourless (e.g. pink or yellow)").

Moreover, the device's instructions for use indicate to check whether the specimen is actually suitable for subsequent analysis following treatment with SLSolution™ device (ref. EIFU013, section instructions for use, clause 6: "Homogenize the specimen by vortexing the test tube for at least 3 seconds at 2000-2500 rpm, checking that the specimen is actually suitable for seeding."). the reason for the recall is that in October 2025, following an increase in notifications received from the market on detected yellow/pink devices, related to the period August-September 2025, Copan verified that the occurrence of device failure (pink/yellow discoloration of reagent) of lots already placed on the market of items 0U019N and 0U020N.A may be unexpectedly high.

Should the defective devices be used, Copan cannot exclude that the functionality in fluidification of respiratory specimens of defective devices is potentially affected.

Hence, in a precautionary manner and considering the unexpectedly high defectivity that may affect marketed lots of items 0U019N and 0U020N.A, Copan is proactively and voluntarily initiating a recall on these devices.

What to do

Contact the manufacturer if you require additional information.

The official notice

This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.

Read the official notice (opens in a new tab)

A little guidance

All guides

Search by name, description or tag. Press Enter to see every match.