Spacelabs Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor (specific units identified by the listed serial numbers and UDI-DI 10841522106415, 10841522106422) is affected. A circuit board issue can cause Cardiac Output measurements to activate early in Auto mode, producing Bad Curve or Irregular Curve errors that may yield inaccurate or missing CO readings or require additional readings, potentially leading to delayed care and/or fluid overload.
What's being recalled
Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Recalling firm: Spacelabs Healthcare, Inc., Snoqualmie, WA, United States.
How to identify it
Serial Numbers: UDI-DI: 10841522106415, 10841522106422. 1496-223736, 1496-223737, 1496-223738, 1496-224375, 1496-224376, 1496-224400, 1496-224401, 1496-224402, 1496-224403, 1496-224404, 1496-224405, 1496-224406, 1496-223850, 1496-223851, 1496-223852, 1496-224448, 1496-224956, 1496-224957, 1496-224958, 1496-224959, 1496-224960, 1496-224961, 1496-224981, 1496-224258, 1496-223595, 1496-223596, 1496-223597, 1496-223598, 1496-223619, 1496-223620, 1496-223621, 1496-223622, 1496-223623, 1496-223624, 1496-224717, 1496-224718, 1496-223854, 1496-224373, 1496-224374, 1496-223894, 1496-223895, 1496-223896, 1496-223897, 1496-223898, 1496-223899, 1496-223900, 1496-223901, 1496-223902, 1496-223903, 1496-223904, 1496-223905, 1496-223906, 1496-223907, 1496-223918, 1496-225043, 1496-225044, 1496-225045, 1496-225046, 1496-225047, 1496-225048, 1496-225049, 1496-225050, 1496-225051, 1496-225052, 1496-225053, 1496-225054, 1496-225055, 1496-225056, 1496-225057, 1496-225058, 1496-225059, 1496-225060, 1496-225061, 1496-225062, 1496-225063, 1496-225064, 1496-225065, 1496-225066, 1496-225067, 1496-225068, 1496-224949, 1496-223760, 1496-223761, 1496-225154, 1496-224639, 1496-224719, 1496-224720, 1496-224721, 1496-225088, 1496-225089, 1496-224672, 1496-223908, 1496-223909, 1496-223910, 1496-223911, 1496-223912, 1496-223913, 1496-223914, 1496-223915, 1496-223916, 1496-223917, 1496-224637, 1496-224638, 1496-224752, 1496-224744, 1496-224745, 1496-224746, 1496-224747, 1496-224407, 1496-224408, 1496-224409, 1496-224410, 1496-224411, 1496-224412, 1496-224413, 1496-224414, 1496-224415, 1496-224416, 1496-224417, 1496-224418, 1496-224419, 1496-224420, 1496-224421, 1496-224753, 1496-224754, 1496-224755, 1496-224756, 1496-224757, 1496-225129, 1496-223825, 1496-223826, 1496-223829, 1496-223830, 1496-223831, 1496-223832, 1496-223833, 1496-223834, 1496-223835, 1496-223836, 1496-223837, 1496-223838, 1496-223839, 1496-222928, 1496-222929, 1496-223774, 1496-223775, 1496-224063, 1496-224064, 1496-223580…
The hazard
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
What to do
Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.
The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.
Where it was sold
Worldwide - US Nationwide distribution including in the states of NY, NJ, AL, OK, IA, MT, WY, MN, TX, WI, PA, WA, OR, MI, CA, SC, ID, FL, MS, KS, SD, LA, OH, MO, NH, GA, MD, TN, AZ, PR and the countries of TAIWAN, CANADA, KUWAIT, FRANCE, SAUDI ARABIA, POLAND, PAKISTAN, MEXICO, ARGENTINA, PANAMA.
The official notice
This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.
Read the official notice (opens in a new tab)