The recall affects the EXENT Analyser (Model IE800.A, serial number 101150, software version v1.0.20, UDI 05051700020886) used with EXENT assays. Plates 2 and 3 were reported as marked “passed” even though it appeared quality control had not been performed on those plates while samples were measured. The potential clinical hazards include false or missed detection of M‑protein and falsely elevated or decreased M‑protein concentrations.
What's being recalled
Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.
Recalling firm: The Binding Site Group, Ltd., Birmingham, N/A, United Kingdom.
How to identify it
Lot Code: Model No; IE800A; EXENT Analyser; UDI; 05051700020886, Serial Number; 101150, Software version; v1.0.20.
The hazard
It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.
What to do
Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.
The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.
Where it was sold
US Nationwide distribution.
The official notice
This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.
Read the official notice (opens in a new tab)