Certain United Orthopedic brand knee trial instruments and femoral valgus adaptors are affected, including U2 Total Knee System PSA Box Trial L and R (catalog 9323-5511–5516 and 9323-5521–5526), USTAR II Hinge Box Trial L and R (9325-5711–5716 and 9325-5721–5726), and specified valgus adaptor/catalog numbers 9323-5333–5340. They were recalled because an incomplete screw process by a subcontractor can prevent proper assembly of the stem trial; the devices were distributed in the U.S., including Texas and Michigan.
What's being recalled
Brand Name: U2 Total Knee System Product Name: PSA Box Trial, L, #1-6 and PSA Box Trial, R, #1-6 Model/Catalog Number: 9323-5511 thru 9323-5516 and 9323-5521 thru 9323-5526 Software Version: N/A Product Description: Knee femur prosthesis trial, reusable Component: No, N/A
Brand Name: USTAR II Knee System Product Name: Hinge Box Trial, L, #1-6 and Hinge Box Trial, R, #1-6 Model/Catalog Number: 9325-5711 thru 9325-5716 and 9325-5721 thru 9325-5726 Software Version: N/A Product Description: Knee femur prosthesis trial, reusable Component: No, N/A
Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and right Model/Catalog Number: 9323-5333 and 9323-5334 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No
Brand Name: U2 Total Knee System Product Name: Femoral valgus offset adaptor, 2mm, 4mm, 6mm, left and Femoral valgus offset adaptor, 2mm, 4mm, 6mm, right Model/Catalog Number: 9323-5335, 9323-5337, 9323-5339 and 9323-5336, 9323-5338, 9323-5340 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No
Recalling firm: United Orthopedic Corp., Kaohsiung, Taiwan.
How to identify it
Product Number; UDI-DI Code; Lot Code: Product Number 9323-5511; UDI-DI 04719886906929; Lot number 25FB54 Product Number 9323-5512; UDI-DI 04719886906936; Lot number 25FB55 Product Number 9323-5513; UDI-DI 04719886906943; Lot number 25FB56 Product Number 9323-5514; UDI-DI 04719886906950; Lot number 25FB57 Product Number 9323-5515; UDI-DI 04719886906967; Lot number 25FB58 Product Number 9323-5516; UDI-DI 04719886906424; Lot number 25FB59 Product Number 9323-5521; UDI-DI 04719886906974; Lot number 25FB29 Product Number 9323-5522; UDI-DI 04719886906981; Lot number 25FB30 Product Number 9323-5523; UDI-DI 04719886906998; Lot number 25FB31 Product Number 9323-5524; UDI-DI 04719886907001; Lot number 25FB32 Product Number 9323-5525; UDI-DI 04719886907018; Lot number 25FB33 Product Number 9323-5526; UDI-DI 04719886907025; Lot number 25FB34 Product Number; UDI-DI Code; Lot Code: Product number 9325-5711; UDI-DI 04719886907032; Lot number 25FB41 Product number 9325-5712; UDI-DI 04719886907049; Lot number 25FB42 Product number 9325-5713; UDI-DI 04719886907056; Lot number 25FB43 Product number 9325-5714; UDI-DI 04719886907063; Lot number 25FB44 Product number 9325-5715; UDI-DI 04719886907070; Lot number 25FB45 Product number 9325-5716; UDI-DI 04719886907087; Lot number 25FB46 Product number 9325-5721; UDI-DI 04719886907094; Lot number 25FB35 Product number 9325-5722; UDI-DI 04719886907100; Lot number 25FB36 Product number 9325-5723; UDI-DI 04719886907117; Lot number 25FB37 Product number 9325-5724; UDI-DI 04719886907124; Lot number 25FB38 Product number 9325-5725; UDI-DI 04719886907131; Lot number 25FB39 Product number 9325-5726; UDI-DI 04719886907148; Lot number 25FB40 Product Number; UDI-DI Code; Lot Code: Product number 9323-5333; UDI-DI 04719886908084; Lot number 25FB50 Product number 9323-5334; UDI-DI 04719886908091; Lot number 25FB51 Product Number; UDI-DI Code; Lot Code: Product number 9323-5335; UDI-DI 04719886906554; Lot number 25FB64 Product number 9323-5337; UDI-DI…
The hazard
Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.
What to do
Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.
The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.
Where it was sold
US Nationwide distribution in the states of Texas and Michigan.
The official notice
This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.
Read the official notice (opens in a new tab)