Skip to content

Health & beauty Canada Health product recall

V-Loc™ 90 and 180 Suture Products

V‑Loc 90 and 180 absorbable wound closure/suture products (models VLOCL0014, VLOCL0316, VLOCM0003, VLOCL2246, VLOCL0306, VLOCL0604, VLOCL0804; lots A6A0972VFY, A5M1060VY, A5M1022VFY, A6A0998VY, A6A0996VY, A5M1180VY, A5M1181VY, A5M1064VY) are affected. A manufacturing quality inspection found a hole in the breather pouch on the Tyvek side (the sterile barrier); investigation determined the briller plate on the sealing station may be causing the hole and released lots may contain affected units. Contact the manufacturer if you require additional information.

What's being recalled

Companies: Covidien Llc, 15 Hampshire Street, Mansfield, Massachusetts, United States, 02048.

How to identify it

Affected products: V-Loc 180 Wound Closure Device Needle P-12; Lot or serial number: A6A0972VFY; Model or catalogue number: VLOCL0014

Affected products: V-Loc 180 Abs Close Device GS-21; Lot or serial number: A5M1060VY; Model or catalogue number: VLOCL0316

Affected products: V-Loc 90 Absorbable Wound Closure Device P-12; Lot or serial number: A5M1022VFY; Model or catalogue number: VLOCM0003

Affected products: V-Loc 180 Absorbable Wound Closure Device - Green Color; Lot or serial number: A6A0998VY; Model or catalogue number: VLOCL2246

Affected products: V-Loc 180 Absorbable Wound Closure Device - Green Color; Lot or serial number: A6A0996VY; Model or catalogue number: VLOCL0306

Affected products: V-Loc 180 Wound Closure Device Needle V-20; Lot or serial number: A5M1180VY; Model or catalogue number: VLOCL0604

Affected products: V-Loc 180 Wound Closure Device Needle V-20; Lot or serial number: A5M1181VY; Model or catalogue number: VLOCL0604

Affected products: V-Loc 180 Wound Closure Device Needle CV-23; Lot or serial number: A5M1064VY; Model or catalogue number: VLOCL0804

The hazard

On 29-Jan-2026 during a quality inspection at a manufacturing site, a V-Loc 180 unit was found with a hole in the breather pouch on the Tyvek side. The breather pouch serves as the sterile barrier system for the suture. Investigation determined the briller plate on the sealing station may be causing the hole. No complaints have been identified for this issue, but lots within the issue bounding have been released to Distribution Centers (DCs) and sold to customers. These released lots may also contain units with holes in the Tyvek side of the breather pouch.

What to do

Contact the manufacturer if you require additional information.

The official notice

This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.