The affected products are Zimmer® Air Dermatome (model 00-8801-001-00) and Zimmer® Dermatome AN handpiece (model 88710100) manufactured between March 2023 and April 2025; lot/serial numbers exceed 10 digits — contact the manufacturer for specific numbers. They were recalled after multiple complaints, including five in Canada, that devices were cutting beyond the proper depth due to a misaligned thickness control bar discovered on June 18, 2025; contact the manufacturer for additional information.
What's being recalled
Companies: Zimmer Surgical, Inc. Also Trading As Relign Corporation, 200 W Ohio Avenue, Dover, Ohio, United States, 44622.
How to identify it
Affected products: Zimmer® Dermatome AN Handpiece; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 88710100
Affected products: Zimmer® Air Dermatome; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 00-8801-001-00
The hazard
Zimmer Biomet is conducting a phased serial specific medical device recall for the Zimmer® Air Dermatome and Zimmer® Dermatome AN handpieces. Multiple complaints have been received related to dermatome devices that were cutting beyond the proper depth. Five complaints have been reported in Canada, they are CMP-0960689, CMP-0969102, CMP-0973010, CMP-0984249 and CMP-0987024. Inspections of the returned complaint devices confirmed that affected devices may have a misaligned thickness control bar. If the device has a misaligned thickness control bar from manufacture, it would be expected to manifest such that any potential negative outcomes would be present upon the first few uses. All devices manufactured between March 2023 and April 2025, that have not been returned for repair or preventive maintenance, are within scope of the recall. The defectiveness was discovered on June 18, 2025. Issue was identified through the complaints handling process.
What to do
Contact the manufacturer if you require additional information.
The official notice
This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.