Azurion and Allura Xper Systems

Philips Azurion and Allura Xper systems (models including Azurion 7 B20/7 M12/7 M20/7 B12 and Allura Xper FD10/FD20/FD20 OR Table—catalog numbers listed in the notice) with X‑ray tube cooling units CU3101 manufactured between February 2016 and May 2020 are affected. The deaeration hose may degrade over time, potentially causing oil leakage that can reduce tube cooling, force the system into low‑dose fluoroscopy with the message “low load fluoroscopy flavor selected: tube cooler problem,” and in rare cases produce fumes that may trigger environmental detectors. Contact the manufacturer for additional information.

What's being recalled

Companies: Philips Medical Systems Nederland B.V., Veenpluis 6, Best, Netherlands, 5684 PC.

How to identify it

Affected products: Azurion 7 B20; Lot or serial number: Not applicable.; Model or catalogue number: 722 068

Affected products: Allura Xper FD10; Lot or serial number: Not applicable.; Model or catalogue number: ALLURA XPER FD10

Affected products: Allura Xper FD10; Lot or serial number: Not applicable.; Model or catalogue number: 001443

Affected products: Allura Xper FD20 (Product Of Allura Clarity Family); Lot or serial number: Not applicable.; Model or catalogue number: 722028

Affected products: Azurion 7 M12; Lot or serial number: Not applicable.; Model or catalogue number: 722 078

Affected products: Azurion 7 M20; Lot or serial number: Not applicable.; Model or catalogue number: 722 079

Affected products: Allura Xper FD10 (Product Of Alluraclarity Family); Lot or serial number: Not applicable.; Model or catalogue number: 722026

Affected products: Allura Xper FD20/10 (Product Of Alluraclarity Family); Lot or serial number: Not applicable.; Model or catalogue number: 722029

Affected products: Allura Xper FD20 OR Table (Product Of Alluraclarity Family); Lot or serial number: Not applicable.; Model or catalogue number: 722035

Affected products: Azurion 7 B12; Lot or serial number: Not applicable.; Model or catalogue number: 722 067

The hazard

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the x-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message "low load fluoroscopy flavor selected: tube cooler problem". This condition cannot be resolved by restarting the system. Under normal conditions, any leaked oil is contained within the system. In rare cases of more significant leakage, oil may extend beyond the intended containment area. If oil then comes into contact with hot surfaces, fumes may be generated. these fumes may activate environmental detection systems.

What to do

Contact the manufacturer if you require additional information.