What's being recalled

Companies: GE Medical Systems Information Technologies Inc., 3114 N Grandview Blvd, Waukesha, Wisconsin, United States, 53188.

How to identify it

Affected products: CareIntellect Perinatal; Lot or serial number: Version 2.3.1; Model or catalogue number: K2504AA

The hazard

GE Healthcare has become aware of an issue in CareIntellect Perinatal (CIP) that may result in loss of monitoring information and clearance of active patient monitoring alarms.

When this issue occurs due to an interruption in system connectivity, monitoring data will be unavailable for up to approximately 20 seconds and currently active visual and audible alarm indicators will no longer be displayed or announced at the CIP user interface. Following resolution of system connectivity, alarm triggers will restart based on the configured thresholds. This issue could delay recognition of a maternal or fetal condition requiring clinical assessment or intervention.

What to do

Contact the manufacturer if you require additional information.

The official notice

This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.