CareLink™ SmartSync™ Device Manager

What's being recalled

Companies: Medtronic Inc., 710 Medtronic Parkway N.E., Minneapolis, Minnesota, United States, 55432.

How to identify it

Affected products: RF HEAD 24967 PATIENT CONNECTOR; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 24967

Affected products: CareLink SmartSync Device Manager App; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: M01A02

Affected products: PROGRAMMER 24970A; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 24970A

The hazard

A complaint from the CareLink network reported a sudden decrease in the Remaining Longevity of a patient's Micra VR CIED from 3 years to 6 months following a remote transmission, which was later found to be incorrect. Analysis revealed a calculation issue in SmartSync and CareLink for Micra VR2 and Micra AV2 devices, occurring under specific conditions. This miscalculation stems from a software error where a function received an integer instead of a real data type, defaulting the average pacing rate to 60 BPM and resulting in inaccurate Remaining Longevity estimates.

What to do

Contact the manufacturer if you require additional information.