What's being recalled
Companies: Oxynov, Inc., 725, Boulevard Lebourgneuf, Quebec, Quebec, Canada, G2J 0C4.
How to identify it
Affected products: FreeO2 Automated Oxygen Therapy Device Adult & Pediatric; Lot or serial number: Version HW 2.3; Model or catalogue number: FO2-110-00
The hazard
Under specific conditions following a deep battery discharge, error 29515 may occur during startup of a FreeO2 HW 2.3 device using software package 1.4.2 and prevent its use. The update to software package 1.4.3 corrects this issue and restores the expected startup behaviour.
What to do
Contact the manufacturer if you require additional information.
The official notice
This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.
Read the official notice (opens in a new tab)