What's being recalled
Companies: Oxynov, Inc., 725, Boulevard Lebourgneuf, Quebec, Quebec, Canada, G2J 0C4.
How to identify it
Affected products: FreeO2 Automated Oxygen Therapy Device Adult & Pediatric; Lot or serial number: Version HW 1.2; Model or catalogue number: FO2-110-00
The hazard
A battery reading error may occur when certain Freeo2 HW 1.2 devices are restarted following an upgrade from software version 1.2.X to 1.4.X. The error is immediately detectable, prevents the device from being returned to service and requires installation of a resistor at position R39.
What to do
Contact the manufacturer if you require additional information.
The official notice
This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.
Read the official notice (opens in a new tab)