Skip to content

Health & beauty Canada Health product recall

Ingenia and Achieva Systems

Philips Medical Systems Nederland B.V. has identified affected MRI systems: Ingenia 1.5T (serial 70851), Ingenia Elition X (serials 45052, 45590, 62158), Ingenia Ambition X (serial 49004), and Achieva 3.0T MRI systems (serials 17385, 38109). The reconstruction voxel size settings in the default MRE scan protocol can be too small when used with Resoundant’s algorithm, which can potentially lead to inaccurate (lower) liver stiffness measurements. Contact the manufacturer if you require additional information.

What's being recalled

Companies: Philips Medical Systems Nederland B.V., Veenpluis 6, Best, Netherlands, 5684 PC.

How to identify it

Affected products: Ingenia 1.5T; Lot or serial number: 70851; Model or catalogue number: 1.5T

Affected products: Ingenia Elition X; Lot or serial number: 45052; Model or catalogue number: 781358

Affected products: Ingenia Elition X; Lot or serial number: 45590; Model or catalogue number: 782107

Affected products: Ingenia Elition X; Lot or serial number: 62158; Model or catalogue number: 782107

Affected products: Ingenia Ambition X; Lot or serial number: 49004; Model or catalogue number: 782138

Affected products: Achieva 3.0T MRI System - Main Unit; Lot or serial number: 17385; Model or catalogue number: ACHIEVA 3.0T

Affected products: Achieva 3.0T MRI System - Main Unit; Lot or serial number: 38109; Model or catalogue number: ACHIEVA 3.0T

The hazard

Philips has become aware through the MRE software system developer Resoundant about an issue affecting MRE stiffness measurements when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm. Specifically, the reconstruction voxel size settings in the default MRE scan protocol are too small, which can potentially lead to inaccurate liver stiffness measurements (lower values).

What to do

Contact the manufacturer if you require additional information.