MAGNETOM Prisma, Vida, Skyra, Verio Systems

What's being recalled

Companies: Siemens Healthcare GmbH, Henkestr. 127, Erlangen, Germany, 91052.

How to identify it

Affected products: MAGNETOM Prisma; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 10849582

Affected products: MAGNETOM Vida; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 11060815

Affected products: Tim 4G - Skyra Fit; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 10849580

Affected products: MAGNETOM Verio, System; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 10276755

Affected products: MAGNETOM Lumina; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 11344916

Affected products: Biograph mMR System; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 10433372

Affected products: MAGNETOM Skyra; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 10432915

Affected products: MAGNETOM Skyra Fit; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 11516217

Affected products: MAGNETOM Cima.X; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 11647158

Affected products: MAGNETOM Vida Fit; Lot or serial number: More than 10 numbers, contact manufacturer.; Model or catalogue number: 11410481

The hazard

This field safety corrective action addresses a potential safety issue associated with the magnet venting paths of some 3 Tesla MRI systems, which will require on-site inspection. An ice blockage may have built up within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up in the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Recall start date: August 14, 2025

What to do

Contact the manufacturer if you require additional information.