Philips Azurion Systems running software version R3.0 are affected; affected units carry the Azurion label descriptions (Azurion 3 M12, 3 M15, 5 M12, 5 M20, 7 B12, 7 B20, 7 M12, 7 M20) and the listed model numbers (for example 722229–722236) and UDIs. They were recalled due to nine identified software issues that may cause loss of X‑ray after Pedal Tap or Phase Fault, missing live X‑ray on the FlexVision monitor, continuous restart or restart after start‑up, AMD triple drive, C‑Partition running out of free space, misalignment of Marker Tool Overlay, or longitudinal position error.
What's being recalled
Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Number: 722229. 2. Azurion 3 M15. Model Number: 722230. 3. Azurion 5 M12. Model Number: 722231. 4. Azurion 5 M20. Model Number: 722232. 5. Azurion 7 B12. Model Number: 722235. 6. Azurion 7 B20. Model Number: 722236. 7. Azurion 7 M12. Model Number: 722233. 8. Azurion 7 M20. Model Number: 722234.
Recalling firm: Philips Medical Systems Nederland B.V., Best, Netherlands.
How to identify it
Software Version: R3.0. 1. Model Number: 722229. UDI Numbers: (01)00884838116726(21)22, (01)00884838116726(21)15, (01)00884838116726(21)16, (01)00884838116726(21)6, (01)00884838116726(21)8, (01)00884838116726(21)4, (01)00884838116726(21)24, (01)00884838116726(21)26, (01)00884838116726(21)20, (01)00884838116726(21)25, (01)00884838116726(21)11, (01)00884838116726(21)21, (01)00884838116726(21)13, (01)00884838116726(21)7, (01)00884838116726(21)10, (01)00884838116726(21)2, (01)00884838116726(21)3, (01)00884838116726(21)29. 2. Model Number: 722230. UDI Numbers: (01)00884838116733(21)2, (01)00884838116733(21)40, (01)00884838116733(21)9, (01)00884838116733(21)1, (01)00884838116733(21)38, (01)00884838116733(21)10, (01)00884838116733(21)45, (01)00884838116733(21)7, (01)00884838116733(21)27, (01)00884838116733(21)28, (01)00884838116733(21)39, (01)00884838116733(21)4, (01)00884838116733(21)5, (01)00884838116733(21)23, (01)00884838116733(21)41, (01)00884838116733(21)20, (01)00884838116733(21)19, (01)00884838116733(21)13, (01)00884838116733(21)15, (01)00884838116733(21)12, (01)00884838116733(21)22. 3. Model Number: 722231. UDI Numbers: (01)00884838116740(21)13, (01)00884838116740(21)29, (01)00884838116740(21)50, (01)00884838116740(21)21, (01)00884838116740(21)19, (01)00884838116740(21)12, (01)00884838116740(21)25, (01)00884838116740(21)4, (01)00884838116740(21)1, (01)00884838116740(21)5, (01)00884838116740(21)2, (01)00884838116740(21)73, (01)00884838116740(21)72, (01)00884838116740(21)75, (01)00884838116740(21)33, (01)00884838116740(21)47, (01)00884838116740(21)60, (01)00884838116740(21)53, (01)00884838116740(21)45, (01)00884838116740(21)61, (01)00884838116740(21)34, (01)00884838116740(21)54, (01)00884838116740(21)35, (01)00884838116740(21)49, (01)00884838116740(21)32, (01)00884838116740(21)28, (01)00884838116740(21)27, (01)00884838116740(21)26, (01)00884838116740(21)51, (01)00884838116740(21)11, (01)00884838116740(21)20, (01)00884838116740(21)41, (01)00884838116740(21)52…
The hazard
Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
What to do
Don't stop taking a prescribed medicine without talking to your doctor or pharmacist first. For other products, stop using them and ask the pharmacist, seller or maker what to do.
The notice doesn't give instructions for consumers, so this is general advice for health & beauty recalls. Check the official notice for the details.
Where it was sold
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Chile, Colombia, Cote de'Ivore, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Kosovo, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Palestine, Philippines, Poland, Portugal, R¿union, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.
The official notice
This page is a summary of the notice published by U.S. Food and Drug Administration (enforcement reports), collected from U.S. Food and Drug Administration (enforcement reports). Check the original before you act on it.
Read the official notice (opens in a new tab)