Spacelabs display cables for Xprezzon® or Qube Bedside Monitors

DVI display cables used with Spacelabs Xprezzon display (model 94267), Xprezzon bedside monitor (model 91393) or Qube compact monitor (model 91390) with lot/serial number 012-089X-XX are affected. During 60601-1-2 testing it was found these cables radiate emissions outside approved limits. Contact the manufacturer, Spacelabs Healthcare Inc., for more information.

What's being recalled

Companies: Spacelabs Healthcare Inc., 35301 Se Center Street, Snoqualmie, Washington, United States, 98065.

How to identify it

Affected products: DVI Cable Used With Medical Grade Flat Panel Display With Touchscreen; Lot or serial number: 012-089X-XX (cable); Model or catalogue number: 94267

Affected products: DVI Cables Used With Qube Compact Monitor; Lot or serial number: 012-089X-XX (cable); Model or catalogue number: 91390

Affected products: DVI Cable Used With Bedside Monitor; Lot or serial number: 012-089X-XX (cable); Model or catalogue number: 91393

The hazard

While conducting 60601-1-2 testing for ongoing product development, it was found that the display cables used to connect an Xprezzon monitor(91393) or Qube compact monitor(91390) to an Xprezzon display(94267) radiates emissions outside the approved limits.

What to do

Contact the manufacturer if you require additional information.

The official notice

This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.

Read the official notice (opens in a new tab)