DVI display cables used with Spacelabs Xprezzon display (model 94267), Xprezzon bedside monitor (model 91393) or Qube compact monitor (model 91390) with lot/serial number 012-089X-XX are affected. During 60601-1-2 testing it was found these cables radiate emissions outside approved limits. Contact the manufacturer, Spacelabs Healthcare Inc., for more information.
What's being recalled
Companies: Spacelabs Healthcare Inc., 35301 Se Center Street, Snoqualmie, Washington, United States, 98065.
How to identify it
Affected products: DVI Cable Used With Medical Grade Flat Panel Display With Touchscreen; Lot or serial number: 012-089X-XX (cable); Model or catalogue number: 94267
Affected products: DVI Cables Used With Qube Compact Monitor; Lot or serial number: 012-089X-XX (cable); Model or catalogue number: 91390
Affected products: DVI Cable Used With Bedside Monitor; Lot or serial number: 012-089X-XX (cable); Model or catalogue number: 91393
The hazard
While conducting 60601-1-2 testing for ongoing product development, it was found that the display cables used to connect an Xprezzon monitor(91393) or Qube compact monitor(91390) to an Xprezzon display(94267) radiates emissions outside the approved limits.
What to do
Contact the manufacturer if you require additional information.
The official notice
This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.
Read the official notice (opens in a new tab)