Spacelabs Healthcare Xhibit® Central Station model 96102 is affected; the issue involves use of the previously distributed splitter (part number 010-1975-00) or customers' own unqualified splitters. These splitters can degrade or completely disable audible and visible alarms. Contact the manufacturer for additional information.
What's being recalled
Companies: Spacelabs Healthcare Inc., 35301 Se Center Street, Snoqualmie, Washington, United States, 98065.
How to identify it
Affected products: Spacelabs Healthcare Xhibit® Central Station; Lot or serial number: Not applicable.; Model or catalogue number: 96102
The hazard
Spacelabs Healthcare has concluded that the use of the previously distributed splitter (part number 010-1975-00) can have an adverse effect on product performance, specifically around audible and visible alarms. Splitters may fail to perform as intended and result in degraded or complete loss of alarm audio and video. Additionally, customer's own use of unqualified splitters may pose the same risks.
What to do
Contact the manufacturer if you require additional information.
The official notice
This page is a summary of the notice published by Health Canada, collected from Health Canada, CFIA and Transport Canada. Check the original before you act on it.